LS 6000 Scintillation System User’s Manual
PN 247971-F
5-20
Data Calculation
Single Photon Monitoring
NOTE Do not leave an empty space for each missing replicate. One empty space indicates to the system that
a new group of replicates follows.
Results for Xtal SL DPM % of Reference
To count the samples, refer to Section 3.2. A typical printout for Xtal SL DPM % of Reference
in Standard Format is shown in Figure 5.17. The calculated result is shown under the % Ref
column heading.
5.9 Single Photon Monitoring
The Single Photon Monitor is used for bioluminescence, chemlluminescence, or other studies
where an accumulated total number of events seen by each photomultiplier tube during each
period of counting is obtained and averaged. Therefore, during single photon monitoring, the
coincidence circuit used for counting radioactivity is disabled.
Setting Up Single Photon Monitor
When Single Photon is selected as Isotope 1, the Data Calculation Program Is automatically
set to Single Photon. Refer to Isotope 1 Section 4
for more information on how the other User
Program parameters are affected. When the Data Calculation menu is displayed, the prompts
for setting up Single Photon Monitor are displayed. The menu is shown in Figure 5.18.
Figure 5.17 Typical Printout for Xtal DPM % of Reference.