LS 6000 Scintillation System User’s Manual
PN 247971-F
6-7
Isotope/DPM Libraries
The Isotope Library
6
10. With both racks in position in the sample changer, press START. The instrument
performs a calibration and counts each sample for
5 minutes.
11. On completion of the count, the instrument stores data for the new isotope(s) in the
library, including type of scintillator, the name, half-life, and window settings. The data
being stored for each isotope is printed. A typical printout is shown in Figure 6.5
. The
Main Menu is displayed. The isotopes added are now available for use in any User
Program.
Deleting Isotopes from the Isotope Library
When an isotope entered in the library is no longer used, or new window settings for an
isotope are desired, delete the isotope from the library. The five isotopes permanently stored
in the Isotope Library cannot be deleted.
To delete an isotope:
Refer to Accessing the Isotope Library Section 6 for information on how to access the New
Isotope Setup Menu
1. From the New Isotope Setup Menu, with “Scintillator” high-lighted, choose Liquid or
Xtal to specify the portion of the Isotope Library you wish to delete from.
2. Highlight “Delete Isotope from Library”. A list of isotopes for the selected scintillator is
displayed in the Data Entry Window.
3. Highlight the isotope to delete and press DELETE.
4. The isotope is deleted from the Isotope Library. That Isotope is no longer available for
use in any User Program.
Figure 6.5 Printout for Setup of New Isotope.