LS 6000 Scintillation System User’s Manual
PN 247971-F
1-4
Introduction
Specifications
System Setup
System Setup allows you to set system parameters tailored to your laboratory requirements
and to set up count parameters and calibration information for Auto DPM. Refer to Section
2.6 for more information.
System Test
System Test provides a number of routines to verily performance of the mechanics,
electronics and memory of the instrument and to test the printer. This mode is used by your
Authorized Beckman Service Representative for servicing the instrument.
Pre-Installation, Installation and Moving
Prior to installation of the instrument or when you want to move the instrument, ensure that
the space and power requirements can be met. Refer to Appendix b for these requirements.
The instrument must be installed, set up, and initially adjusted by your Authorized Beckman
Service Representative.
CAUTION Operation of the instrument before it has been installed and adjusted by your Authorized Beckman
Service Representative may void the warranty.
If it becomes necessary to move the instrument, be careful to avoid any mechanical shock.
For transporting the instrument to another location, contact your Authorized Beckman
Service Representative.
1.3 Specifications
Specifications for the instrument are provided in Appendix a.
1.4 Precautions and Limitations
The following points are important for accurate and trouble-free operation:
1. The symbol A appears on the instrument. Read this user’s guide thoroughly before
operating the instrument. The user’s guide contains important information about the safe
and proper use of your system.
2. Leave the power on at all times once the instrument is installed. Initial warm-up and
stabilizing time varies from one instrument to another. Leaving the power on ensures
you will not take critical measurements while the instrument is still stabilizing after
powering up.
3. Perform a background count and efficiency check periodically to detect possible
radioactive contamination of the detector system and to check overall performance.
Refer to Section 7.5 for information on performing these checks.
4. There is a limit to the amount of cpm the LS counter can process. If a sample is rejected
because it exceeds this limit, it can be counted by diluting it.
5. Sample vials of the wrong size or shape, or standard vials with overhanging caps, may
jam the internal mechanism. Figure 1.1 provides specifications for vial size and
configuration.
6. Even though the instrument is equipped with a static eliminator, static charges can build
up on vials and cause errors in counting. In particular, handling plastic vials with latex