LS 6000 Scintillation System User’s Manual
PN 247971-F
6-10
Isotope/DPM Libraries
Setting Up A Quench Curve
5. If setting up quench curves for applications other than low level studies, highlight
“Normal Quench Curves”. The quench curves are stored in the Standard DPM Library. If
setting up standards for low level studies, highlight “Low Level Quench Curves”. The
quench curves are stored in the Low Level DPM Library. Press [SELECT]. The Main
Editing Window displays two prompts again: Setup Quench Curves and
Review/Edit/Delete Quench Curves.
6. Highlight “Setup Quench Curves” and press SELECT. The Main Editing Window
displays two prompts: Check Precision of Unquenched Std’s and Setup Quench Curve.
7. Highlight “Check Precision of Unquenched Std’s” and press SELECT. The menu shown
in Figure 6.6 is displayed.