LS 6000 Scintillation System User’s Manual
PN 247971-F
a-2
Instrument Specifications
Power Requirements
a.5 Power Requirements
a.6 Dimensions
a.7 Ambient Temperature Range
15
o
C to 35
o
C. Calibration must be performed within 5
o
C of operating temperature.
a.8 Temperature Control Accessory
Electrical Requirements:
(50/60 Hz) 120V 240V BTU/HR
Instrument 3A 1.5A 1230
Temperature 7.5A 3.8A 3060
Control Accy
Inrush Current: 5A at 120V
2.5A at 240V
Power Failure
Recovery
With a fully charged battery, the system returns to a disrupted Automatic
Count for several weeks.
Width
cm(in.)
Height
cm(in.)
Depth
cm(in.)
Weight
kg(lbs.)
Instrument:
91.5(36) 66(26) 80(31.5) 210(460)
Monitor:
33(13) 34.3(13.5) 35.6(12) 8.2(18)
Printer:
39(15.5) 10(4) 30.5(12) 5(11)
Temperature:
35.5(14) 56(22) 73.7(29) 45.4(100)
Control Accy.*
*Temperature Control Accessory attaches to right side of instrument.
Maximum Relative Humidity: 85%
Ambient Temperature Range:
15
0
to 30
0
Maximum Pull Down Time: 8 Hours
Set Point 1:
12
o
C +3
o
C