LS 6000 Scintillation System User’s Manual
PN 247971-F
3-11
Counting Samples
Multi-Task
3
2. From the Multi-Task Menu, highlight “Interrupt Count” and press SELECT. The menu
shown in Figure 3.9 is displayed.
3. Highlight “Count Single Rack” and press SELECT. The Count Single Rack Menu (Figure
3.3) is displayed.
NOTE If Interrupt Data is stored, this data must be reviewed, printed or deleted before another Interrupt
Count can be performed. Refer to Interrupt Data Section 3 for information on Interrupt Data. Press
[PREVIOUS MENU] to return to the Count Single Rack Menu when you are finished in Interrupt Count.
4. To count using the default parameters without any changes, press SELECT.
To change the counting time, highlight “Counting Time”, and enter the desired counting
time (0.01 to 999.99 minutes). Highlight “Count Any Isotope”, and press SELECT.
To count with a User Program, refer to Counting with a User Program at the end of this
section.
5. The Main Editing Window displays instructions to load the samples. Load the Interrupt
Rack with the samples you wish to count as described on the screen.
NOTE The Interrupt Rack with the Interrupt Card installed must be used to count samples during Multi-Task.
Once the samples are counted, the rack is skipped when Auto Count is resumed.
6. Press START to begin Interrupt Count. The screen displays a message “Suspending Auto
Count” and the sample being counted is terminated. The Sample Rack in position is
moved out of the count position and the instrument searches for the Interrupt Rack. The
screen displays “Searching for Interrupt Rack”.
Figure 3.9 The Interrupt Menu