4 - 10 MP6000 INTEGRATOR GUIDE
The weight checked at each position on the scale, shown in Figure 4-4, represent halfway between the center
of the platter and corner. Use a stack configuration of weight
s if needed (i.e., (2) 5.00 lb weights, or (2) 2.500 kg
weights)
depending on the units of measure programmed.
Figure 4-4
Weight Positions on the Scale Platter
5.000 5.000 .005 kg .003 kg
2.500 2.500 .0025 kg .002 kg
0.100 0.100 .0025 kg .001 kg
Return to Zero Test
0.000 0.000 0.0 kg 0.0 kg
Shift Test (see
Figure 4-4)
5.00 Position 1 5.000 .005 kg .003 kg
5.00 Position 2 5.000 .005 kg .003 kg
5.00 Position 3 5.000 .005 kg .003 kg
5.00 Position. 4 5.000 .005 kg .003 kg
Table 4-5
Verification Tests for Metric Scale (Continued)
Verification Test for
Metric Scale Applied
Load kg
Indication = Applied
Load Within
Applicable
Tolerances
All Tolerances
Are +/-
All Tolerances
Are +/-
Acceptance
Single Int
Acceptance
Dual Int
Scale Platter
Shaded Area
Represents the
Horizontal Window