4 - 4 MP6000 INTEGRATOR GUIDE
Legal Scale Dampening Filter
Set the scale sensitivity to vibration by scanning the appropriate bar code (see page 4-17). To program this
parameter, the scale must be in
Calibration Mode. Bar code command values are as follows:
0 = Higher (most sensitive to vibration)
1 = Low (factory default)
2 = Very Low
3 = Ultra Low
The lower the number, the more sensitive the scale is to vibr
ation. The higher the number, the slower the scale
may operate.
Step 3 - Calibration at NO LOAD
When Calibration Mode enters properly, the Scale Display blinks between CAL00 and CAL _ _ and the
7-segment display scrolls
C00Lb, or C00g. Scale calibration can begin.
1. Ensure the platter is installed and there is NO WEIGHTED LOAD on it.
2. Touch the Scale Zero button on the front panel.
3. If Calibration at NO WEIGHTED LOAD is successful after the Scale Zero button is pressed, the process
continues with Step 4 - Calibration at LOAD on page 4-4.
or
If Calibration at NO WEIGHTED LOAD is not successful after the Scale Zero button is pressed, the
process continues with Calibration Failure on page 4-7.
Step 4 - Calibration at LOAD
The Scale Display blinks either CAL25 or CAL11, depending on the units of measure programmed
(CAL25 = pounds; CAL11 = kilograms).
The diagnostic 7 -segment display scrolls C25Lb or C11g.
To continue with calibration:
IMPORTANT The steps in Step 4 - Calibration at LOAD must be performed within 90 seconds after pressing the
Scale Zero button.
IMPORTANT These steps must be performed within 90 seconds after pressing the Scale Zero button in Step 3 -
Calibration at NO LOAD.